Ali, Syed Mastan and Bharath, Ponnuri and Ramakrishna, V. Siva and Ramachandran, D. (2022) Quantification of Impurity-G Content in Levetiracetam Injection (100 mg/ml) by Using RP-HPLC Technique. Journal of Pharmaceutical Research International, 34 (28B). pp. 53-61. ISSN 2456-9119
6176-Article Text-8337-1-10-20221006.pdf - Published Version
Download (293kB)
Abstract
Extremely responsive technique for the quantification of impurity-G content in Levetiracetam samples by RP-HPLC method has been presented in the here manuscript. Quantification of impurity-G content in Levetiracetam sample by HPLC with UV Detector. Impurity-G was concluded by RP-HPLC technique by waters X-Terra MS-C18 (250 x 4.6mm, 5µm) column as motionless segment. Column oven preserved 45°C and taster cooler temperature 5°C, injection quantity 50µL, flow rate was 1.0 mL/min, impurity-G was detected using UV detector at the wavelength of 200nm and run time was 30 minutes. The movable segment utilized pH 3.0 buffer and acetonitrile in the proportion of (850:150 v/v) isocratic mode. The technique legalization has been carried as per International Conference on Harmonization guidelines (ICH). Limit of quantitation (LOQ) was found 0.4122 µg/mL for impurity-G.
Item Type: | Article |
---|---|
Subjects: | EP Archives > Medical Science |
Depositing User: | Managing Editor |
Date Deposited: | 08 Mar 2023 07:41 |
Last Modified: | 12 Mar 2024 04:05 |
URI: | http://research.send4journal.com/id/eprint/1426 |